Abstract

We have developed an ultra-low field (ULF) NMR/MRI system using an HTS-rf-SQUID and evaluated performance of the system as a contaminant detection system for foods and drinks. In this work, we measured 1D MRIs from water samples with or without various contaminants, such as aluminum and glass balls using the system. In the 1D MRIs, changes of the MRI spectra were detected, corresponding to positions of the contaminants. We measured 2D MRIs from food samples with and without a hole. In the 2D MRIs, the hole position in the sample was well visualized. These results show that the feasibility of the system to detect and localize contaminants in foods and drinks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.