Abstract

The patients in NICU (Neonatal Intensive Care Unit) are more likely to get portable X-ray often while they are in the hospital. These patients potentially may get relatively more exposure dose in total in a short period of time. We developed a software to analyze the exposure dose for the patients in the incubator, which is called PIETAII (Patient Information of Exposure dose Total Analysis in NICU). Then, we compared the accuracy of PIETAII and SDEC (surface dose evaluation code) based on customary method. Using the 5 cm body thickness with exposure setting of 50 kV, 1 mAs, relative error between the customary method and the calculated dose by PIETAII and SDEC were 1.96% and 32.35%, respectively. PIETAII is a useful software to estimate the entrance surface dose using the exposure setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.