Abstract
The newly developed semiautomatic data acquisition and processing system allows progress in quantitative evaluation of functions of implanted tilting disk valves. In the new system, data acquisition could be carried out semiautomatically by utilizing image processing methods, and the data analysis time was shortened. The system consists of a computer-controlled cinefilm loading device, a CCD camera, a frame memory, a superimposer, and CRT displays. Since series of the open- and shut-mode frames are commonly observed by turns in normal subjects, data acquisition was carried out manually at the first frame of each mode and semiautomatically from the second frame to the last. By processing the image data with the binarization method, the valve contours were detected, and an open angle and other motion properties can be obtained. Analyzing 90 frames of the cinefilm, this system analyzes data in a time period 20 min shorter than the manual procedure. With the implanted Medtronic Hall valve, a 60 to 65 degrees open angle was calculated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.