Abstract

Contamination adhered to the surface of a µCMM stylus tip compromises the measurement accuracy of the µCMM system, potentially causing dimensional errors that are over ten times larger than the uncertainty of a modern µCMM. In prior work by the authors, the use of a high pressure CO2 gas stream was demonstrated to achieve significant cleaning rate for a range of contaminant without damage to the stylus tip surface. This paper explores the practical challenges of achieving effective stylus tip cleaning in situ on µCMM systems. Two types of snow cleaning approaches were evaluated for their coverage of cleaning, thermal impact and gas flow forces. This work then presents a novel multi-nozzle prototype system using pulsed snow streams to achieve cleaning coverage over the entire stylus tip, and balances forces from the snow streams reducing drag force imparted by the gas stream to levels comparable to the probing force of µCMMs, as well as allowing automated cleaning procedure integrated into a µCMM system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.