Abstract

This study reports on the development of CCD-based optical computed tomography (CT) CT-s2. A commercially available 10× fast optical computed tomography scanner (OCTOPUSTM-10X, MGS Research, Inc., Madison, CT, USA) was used for comparison. NIPAM polymer gel dosimeter was used to validate the performance of CT-s2. The gamma pass rate can reach 96.00% when using a 3% dose difference and 3 mm dose-to-agreement criteria. The results of CT-s2 are as good as those of the single-beam optical-CT scanner, but the scanning time of CT-s2 is only one-tenth of that of the single-beam optical-CT scanner.