Abstract

For the absolute measurement of temperature using Radiometer system, Radiometer must be calibrated with known incident power. Calibration setup is indigenously developed to calibrate the various (E, Ka AND F-Band) ECE Radiometer systems available at IPR. "Hot and Cold technique" is used with blackbody sources, one at room temperature and another at Liquid Nitrogen temperature.New calibration technique using DSP, as well as the conventional technique using Lock-in amplifier, are implemented to remove the noise of system. The blackbody sources are chopped with high accuracy DC motor to generate the temperature difference. Calibration coefficients for individual channels are determined. Inherent noise and losses are calculated by analytical way as well as experimental measurements.Theoretical and experimental measurement has shown necessitate for High temperature black body target. Prototype is generated for High temperature blackbody calibration source. Various materials are analyzed for thermal and dielectric properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.