Abstract

We focus on research to develop a compact human blood sampling device used for the Self Monitoring of Blood Glucose (SMBG). The SMBG comprises: (1) an indentation system using a shape memory alloy (SMA) actuator to force a microneedle through the skin; (2) a micro electrical pumping system to extract blood using a bimorph type piezoelectric microactuator; (3) a biosensor using an enzyme such as glucose oxidase (GOx) to detect and evaluate the amount of glucose in extracted blood. A titanium microneedle the same size as a female mosquito's labium (60 mum outer diameter, 25 mum inner diameter) was produced by the sputter deposition method. The mechanical design of the device was based upon the mosquito's blood sampling mechanism. The blood extraction system worked well. We measured the performance of the principal components: the indentation load for a microneedle of external diameter 100 mum was found to be 0.1 N. The pumping system has an extraction speed of about 2 mul/min for whole blood. This is similar to that achieved by the mosquito.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.