Abstract

Based on the analysis of the theoretical calculation model of axial resolution of optical coherence tomograph for the posterior segment of the human eye, a set of testing device for measuring its axial resolution is designed and developed. In view of a commercial ophthalmic optical coherence tomograph in clinical use, its axial resolution is calculated to be 5.07 μm theoretically, and the actual measurement value is 5.45 μm. The uncertainty of the detection device is evaluated and the result is (5.45±0.10) μm. The measurement error introduced by the testing device is very small. Meanwhile, the axial resolution measured by the testing device meets the requirements of the instrument(≤ 6 μm).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call