Abstract
An SPE procedure for the determination of zaleplon and zopiclone in low-volume human hemolyzed blood using fast GC with negative-ion chemical ionization MS has been developed and validated. Both analytes were well retained on Oasis MCX and HLB columns, and sufficient extraction efficiency was achieved at pH 9.0. For further study a hydrophilic-lipophilic sorbent Oasis HLB was selected due to the polarity of sorbent surface and its large surface area in order to achieve efficient extraction of both analytes in a single step. Special attention has been paid to choosing washing and eluting solvents, resulting in a particularly/extremely clean and moisture-free extract. The mean extraction efficiency was higher than 90.1% for zaleplon and 82.9% for zopiclone. The precision for zaleplon and zopiclone was between 3.04-10.58% and 4.08-9.52%, respectively. Whereas the accuracy was in the range from -5.73 to 6.00%, and from -7.00 to 6.32% for zaleplon and zopiclone, respectively. The results show that the developed method is accurate, selective, precise, and very fast with excellent recovery and low LOD and LOQ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.