Abstract
We report on the development of an on-site therapeutic drug monitoring (TDM) method for vancomycin (VCM) utilizing a portable spectrometer and commercially available immunoturbidimetric assay reagents designed for automated clinical chemistry analyzers. The method enables the quantification of VCM in plasma samples within 10min, with a good correlation between the measured values and the theoretical values (r2 = 0.995). The intra and inter-day precisions were found to be below 12.5% and 17.7%, respectively. Moreover, we established a correlation between the quantitative values using this method and those measured through HPLC-UV and automated clinical chemistry analyzers, showing good reliability (R2 = 0.970 and 0.951, respectively). This method allows anyone to rapidly perform TDM at the bedside and is expected to be used to evaluate appropriate drug therapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.