Abstract

We evaluated commercially available, portable, on-site equipment for chemical warfare agent detection (a gas detection tube, ion mobility spectrometer, surface acoustic wavelength detector, flame photometric detector, photoionization detector, Fourier-transformed infrared spectrometer and a gas chromatograph-mass spectrometer) using authentic, vaporized chemical-warfare agents from the standpoint of their qualitative detection characteristics, detection limits, response times, frequency of false alarms and residubility on the devices. False alarms and the strong adsorption of agents by the devices are typical drawbacks of such equipment. As a screening method for biological warfare agents, on-site methods using flow cytometry, bioluminescence assay, and lateral flow immunoassay were developed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.