Abstract
A time-of-flight mass spectrometer (TOF-MS) that incorporates an ion storage trap as a pulsed extraction source and sample preconcentrator has been developed for air monitoring. This new instrument configuration is designed for real-time monitoring of changes in trace volatile and semivolatile compound concentrations in air. Typically, detection limits are in the low parts per billion (v/v) range. This ion store/TOF-MS configuration permits a relatively large volume air flux to be drawn directly into the ion trap cavity, creating a local pressure that exceeds 10-4 Torr. Operating the ion trap at this pressure increases the extent of chemical ionization with the abundant N2+ and O2+ that are formed by electron impact. The air background also serves as a suitable collisional damping matrix, helping to compress the ion cloud to the center of the trap, which is ideal for TOF-MS analysis. Ions are typically accumulated in the trap for 400 ms, at which point they are extracted through the exit end cap and analy...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.