Abstract

The aim of this work is development of methodology for analysis of inorganic cations (sodium, potassium, magnesium, calcium) in seawater used for OTC nasal spray production. It is shown that determined ion chromatographic parameters: type of analytical column, eluent flow rate, concentration of eluent competing ions, and volume of sample loop enables baseline separation of all components of interest. By performing validation procedures and a number of statistical tests, it is shown that a developed ion chromatographic method has superior performance characteristic: linearity R 2 ≥ 0.9990, recovery = 99.88–100.63%, repeatability RSD ≤ 1.78%. That result proves that the proposed method can be used for routine quality assurance analysis in the OTC pharmaceutical industry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.