Abstract
A 256-element mercuric iodide (HgI/sub 2/) detector array has been developed which is intended for use as an intraoperative gamma camera (IOGC). The camera is specifically designed for use in imaging gamma-emitting radiopharmaceuticals (such as 99m-Tc labeled Sestamibi) incorporated into brain tumors in the intraoperative surgical environment. The system is intended to improve the success of tumor removal surgeries by allowing more complete removal of subclinical tumor cells without removal of excessive normal tissue. The use of HgI/sub 2/ detector arrays in this application facilitates construction of an imaging head that is very compact and has a high SNR. The detector is configured as a cross-grid array. Pixel dimensions are 1.25 mm squares separated by 0.25 mm. The overall dimension of the detector is 23.75 mm on a side. The detector thickness is 1 mm which corresponds to over 60% stopping at 140 keV. The array has good uniformity with average energy resolution of 5.2/spl plusmn/2.9% FWHM at 140 keV (best resolution was 1.9% FWHM). Response uniformity (/spl plusmn//spl sigma/) was 7.9%. A study utilizing realistic tumor phantoms (uptake ratio varied from 2:1 to 100:1) in background (1 mCi/l) was conducted. SNRs for the reasonably achievable uptake ratio of 50:1 were 5.61 /spl sigma/ with 1 cm of background depth (normal tissue) and 2.74 /spl sigma/ with 4 cm of background for a 6.3 /spl mu/l tumor phantom (/spl sim/270 nCi at the time of the measurement).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.