Abstract

Objective: The level of primary implant fixation in cementless total hip arthroplasty is a key factor for the longevity of the implant. Vibration-based methods show promise for providing quantitative information to help surgeons monitor implant fixation intraoperatively. A thorough understanding of what is driving these changes in vibrational behavior is important for further development and improvement of these methods. Additionally, an instrument must be designed to enable surgeons to leverage these methods. This study addresses both of these issues. Method: An augmented system approach was used to develop an instrument that improves the sensitivity of the vibrational method and enables the implementation of the necessary excitation and measurement equipment. The augmented system approach took into account the dynamics of the existing bone-implant system and its interaction with the added instrument. Results: Two instrument designs are proposed, accompanied by a convergence-based method to determine the insertion endpoint. The modal strain energy density distribution was shown to affect the vibrational sensitivity to contact changes in certain areas. Conclusion: The augmented system approach led to an instrument design that improved the sensitivity to changes in the proximal region of the combined bone-implant-instrument system. This fact was confirmed both in silico and in vitro. Clinical Impact: The presented method and instruments address practical intraoperative challenges and provide perspective to objectively support the surgeon’s decision-making process, which will ensure optimal patient treatment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call