Abstract

An in situ polarization-dependent total-reflection fluorescence X-ray absorption fine structure (PTRF-XAFS) spectroscopy system has been developed, which enables PTRF-XAFS experiments to be performed in three different orientations at various temperatures (273-600 K) and pressures (10(-10) approximately 760 torr). The system consists of a measurement chamber and a preparation chamber. The measurement chamber has a high-precision six-axis goniometer and a multielement solid-state detector. Using a transfer chamber, also operated under ultra-high-vacuum conditions, the sample can be transferred to the measurement chamber from the preparation chamber, which possesses low-energy electron diffraction, Auger electron spectroscopy and X-ray photoelectron spectroscopy facilities, as well as a sputtering gun and an annealing system. The in situ PTRF-EXAFS for Cu species on TiO2 (110) has been measured in three different orientations, revealing anisotropic growth of Cu under the influence of the TiO2 (110) surface.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.