Abstract

A simple, accurate and sensitive capillary gas chromatographic test procedure was developed for the detection and quantification of impurities in bulk 3-tropanyl-3,5-dichlorobenzoate, a drug which is a neural serotonin 5HT3 receptor antagonist. The drug substance was dissolved in acetonitrile and chromatographed on a 30 m×0.32 mm 0.25 μm film DB-5 column operated with a temperature program of 80 to 230°C. A flame ionization detector was used, and the impurities detected in the drug were estimated from peak areas on a percent basis compared to the parent peak. Validation of this procedure included a recovery study of spiked impurities from 0.1 to 1.2% (w/w) and a repeatability study using two different synthesized batch lots of the drug.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.