Abstract
Field ionization from sharp tips is attracting increased attention for use in detectors for neutral atomic/molecular species. However, the direct detection of the ionized species typically results in low sensitivities due to the small acceptance angle of the receiving ion-sensitive measurement device (usually a channel electron multiplier) and can result in sputtering damage due to the relatively high mass and energy of the incident ion species. Here we present a design for a field-ionization-based neutral atom detector incorporating a simple secondary electron generating stage. The use of such a stage decouples the field-ionized species from the detected electron signal, thus eliminating any sputtering damage to the channel electron multiplier. The detector armature discussed is shown to exhibit a linear response to neutral gas pressure and a sensitivity that is improved by more than two orders of magnitude over a previous field ionization detector design.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.