Abstract

We developed a robot system that can be used for image-guided experimental procedures on small animals, where the goal is to perform physical actions at specific positions identified on a preoperative image. The animal is first placed in a fixture that is compatible with all imaging systems of interest, including PET, SPECT, CT and MRI. After imaging, the fixture is attached and registered to the robot system, where the image-guided intervention is performed. This system has been applied to perform pO2 measurements with physical probes based on tumor hypoxia images obtained in an animal PET scanner. This paper focuses on the design and validation of the robot system. The validation is performed using a phantom and includes a new method for estimating the Fiducial Localization Error (FLE) that is based on the measured Fiducial Distance Error (FDE). The results indicate that the robot system can position the measurement probe at a defined target with a mean error that is less than 0.4 mm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.