Abstract

An indirect solid-phase enzyme immunoassay (EIA) was developed for the detection of poliovirus antigen. Virus antigen was obtained in LLC-MK2 cell cultures and used to prepare antibodies in rabbit and guinea pig. Antibodies were evaluated by double immunodiffusion and neutralization test. Optimal concentrations of guinea pig and rabbit immunoglobulins were determined by checkerboard titration. Microtitre plates were coated with 15.0 µg/ml guinea pig anti-polio immunoglobulin and rabbit anti-polio immunoglobulin at the concentration of 7.94 µg/ml was used as detecting antibody. The standard curve with eight different antigen concentrations in eight replicates resulted in a coefficient of variation (CV) between 2.1% to 7.8%. The dose-response relationship was determined by simple linear regression with a coefficient of correlation (R²) equal to 96.4%. The assay detected a minimum of 2.3 µg/ml poliovirus antigen.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.