Magnetic bead immunocapture-LC-MS has been widely used for bioanalysis of biotherapeutic proteins. However, magnetic beads are difficult to be fully automated and more costly than ELISA plates. Develop an ELISA-LC-MS hybrid assay as an alternate platform. Among seven ELISA plates tested, Pierce streptavidin plates, which did not require time-consuming capture antibody precoating steps, provided the best sensitivity and assay dynamic range (5-2500ng/ml or 10-5000ng/ml), similar to magnetic bead immunocapture-LC-MS assay and better than an ELISA (50-500ng/ml). The entire procedures could be fully automated using a liquid handling system. This study demonstrates that ELISA-LC-MS hybrid approach using streptavidin plates represents a promising platform for bioanalysis of biotherapeutics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.