Abstract

In this paper, the development and application of a multiple heart-cutting achiral-chiral LC-LC method (mLC-LC) for the analysis of dansylated (Dns) branched-chain amino acids in commercial tablets are described. In the first dimension, a Waters Xbridge RP C18 achiral column was used under gradient conditions with buffered aqueous solution and acetonitrile. The elution order Dns-valine (Dns-Val)<Dns-isoleucine (Dns-Ile)<Dns-leucine (Dns-Leu) turned out with full resolution between adjacent peaks: 7.25 and 1.50 for the Val/Ile and the Ile/Leu pairs, respectively. A "research" validation study was performed, revealing high accuracy (Recovery%) and precision (RSD%) using two external set solutions, respectively, in the range 93.7%-104.1% and 0.4%-3.2%. The C18 column was connected via a two-position six-port switching valve to the quinidine-based Chiralpak quinidine-anion-exchange chiral column. A water/acetonitrile, 30/70 (v/v) with 50mM ammonium acetate (apparent pH of 5.5) eluent allowed getting the three enantiomers' pairs resolved: RS equal to 4.3 for Dns-Val and Dns-Ile, and 1.7 for Dns-Leu. The application of the mLC-LC method confirmed that the content of Val, Ile, and Leu in the tablets was compliant with that labeled by the producer. Only l-enantiomers were found in the food supplement, as confirmed by LC-MS/MS analysis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call