We report the development of a dual-mode mass-directed supercritical fluid chromatography and reversed-phase liquid chromatography purification system. The addition of a third pump allows for flexible mobile phase control between the two techniques, and enables operation of either chromatography mode within minutes by activation of a set of switching valves on a single system. Software control, fluidic pathways, interface to the mass spectrometer, and fraction collection have been modified for compatibility between both separation methods. The conditioning solvent and tuning parameters for the mass spectrometer were adjusted to achieve an ideal signal trace in either mode with good linearity (r(2) > 0.970) over a range of concentrations and minimal noise for accurate peak detection and isolation. The registration success rate is 90% and overall sample recovery for either technique is 80-90%. Combining two orthogonal separation and purification modes in one single system has improved the purification throughput of complex mixtures and has been a valuable, cost-saving tool in our laboratory.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.