Abstract

Early diagnosis is essential for the successful management of Burkholderia pseudomallei infection, but it cannot be achieved by the current gold standard culture technique. Therefore, this study aimed to develop a lateral flow immunoassay (LFIA) targeting B. pseudomallei capsular polysaccharide. The development was performed by varying nitrocellulose membrane reaction pads and chase buffers. The prototype LFIA is composed of Unisart CN95 and chase buffer containing tris-base, casein, and Surfactant 10G. The assay showed no cross-reactivity with E. coli, S. aureus, P. aeruginosa, and P. acne. The limit of detections (LODs) of the prototype LFIA was 107 and 106 CFU/mL B. pseudomallei in hemoculture medium and artificial urine, respectively. These LODs suggest that this prototype can detect melioidosis from positive hemoculture bottles but not straight from urine. Additionally, these LODs are still inferior compared to Active Melioidosis Detect (AMDTM). Overall, this prototype holds the potential to be used clinically with hemoculture bottles. However, further improvements should be considered, especially for use with urine samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.