A detector for microchromatography in which Raman spectroscopy is used to identify the eluted species has been developed. The detector is designed to be applicable to a wide range of compounds without requiring the presence of a chromophore. Its use is illustrated in the analysis of nitro compounds on a 250 μm i.d. column. Raman spectra of each of the compounds could be identified as they passed the detector. The advatages of the use of fully deuterated solvents are demonstrated by the analysis of nitrobenzenes in methanol/water mobile phases. The detection limit for nitrobenzene using the Raman line at 1342 cm−1 was 75ng.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.