Abstract

A complex impedance measurement device with a short response time and high noise immunity is presented in this paper. The device based on a radio-frequency reflectometer was specially developed for electro-physical property investigations of materials in quasi-isentropic compression experiments. The maximum operating frequency of the device is up to 600MHz for reducing intense low-frequency noises. Meanwhile, an off-line signal processing code was developed to improve the response time of the device to less than 10ns. Using the device, the complex impedance and electrical conductivity of water compressed by an explosive-driven magnetic flux compression generator were measured, and an abrupt change in the complex impedance of water caused by a liquid-solid transition was directly observed under intense electromagnetic interference.