Abstract
This study focuses on the development of transducers for Osstell® Mentor in measuring stability of various implant systems. Experiments are performed to verify the compatibility of a self-made transducer. An acoustic signal analysis system is introduced to analyze the acoustic signal emitted from the Osstell® Mentor probe, and the frequency with peak amplitude in the amplitude-frequency plot is identified. The resonance frequency obtained is found to be linearly dependent upon the ISQ value by Osstell® Mentor. This finding is helpful for developing compatible transducers for various implant systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.