Abstract

Recently, a robotic system was developed to assist in Single-Port Endoscopic Surgery (SPS). However, the existing system required a manual operation of vision and viewpoint, hindering the surgical task. We proposed a surgical endoscopic robot for SPS with a dynamic vision control, the endoscopic view being manipulated by a master controller. The prototype robot consists of a manipulator for vision control, and dual tool tissue manipulators (gripping: five DOFs; cautery: three DOFs) can be attached at the tip of the sheath manipulator. In particular, this paper focuses on the details of the mechanism and control scheme of the tool manipulator. The experimental results show that our manipulator exhibits a response with a precision of less than 0.15 mm and a time delay of less than 31 ms, when the input frequency is 1.0 Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.