Abstract
Constructing tissue-mimicking phantoms of the brain for ultrasonic studies is complicated by the low backscatter coefficient of brain tissue, causing difficulties in simultaneously matching the backscatter and attenuation properties. In this work, we report on the development of a polyvinyl alcohol-based tissue-mimicking phantom with properties approaching those of human brain tissue. Polyvinyl alcohol was selected as the base material for the phantom as its properties can be varied by freeze–thaw cycling, variations in concentration and the addition of scattering inclusions, allowing some independent control of backscatter and attenuation. The ultrasonic properties (including speed of sound, attenuation and backscatter) were optimized using these methods with talc powder as an additive. It was determined that the ultrasonic properties of the phantom produced in this study are best matched to brain tissue in the frequency range 1–3 MHz, indicating its utility for laboratory ultrasonic studies in this frequency range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Ultrasound in Medicine & Biology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.