Abstract

ObjectiveIn this study, a novel, simple, and rapid immunoassay for the determination of gastrin-17 (G-17) in human serum was established by combining immunomagnetic beads with time-resolved fluorescence immunoassay (TRFIA). MethodsImmunomagnetic beads were coated with anti-G-17 M01 antibody, anti-G-17 M02 antibody was labeled with Eu3+ chelates. The concentration of G-17 in the serum was detected with the double-antibody sandwich method. ResultsThe limit of background(LOB), limit of detection (LOD), and limit of quantification (LOQ) were 0.09, 0.104, and 0.39 pmol/L, respectively. The detection range of G-17-TRFIA was 0.39–100 pmol/L. The average intra- and inter-assay coefficients of variation (CV) were 5.95%–9.07% and 6.09%–8.14%, respectively. The recoveries for the serum samples ranged from 94.70% to 100.95%. The specificity of our G-17-TRFIA was acceptable. The correlation coefficient between G-17-TRFIA and commercial G-17-ELISA methods was R2 = 0.9092. ConclusionsA novel G-17-TRFIA detection method was successfully established to provide a reference for the early diagnosis of patients with atrophic gastritis in clinical research.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.