Abstract

A sound-coding strategy for users of cochlear implants, named enhanced-envelope-encoded tone (eTone), was developed to improve coding of fundamental frequency (F0) in the temporal envelopes of the electrical stimulus signals. It is based on the advanced combinational encoder (ACE) strategy and includes additional processing that explicitly applies F0 modulation to channel envelope signals that contain harmonics of prominent complex tones. Channels that contain only inharmonic signals retain envelopes normally produced by ACE. The strategy incorporates an F0 estimator to determine the frequency of modulation and a harmonic probability estimator to control the amount of modulation enhancement applied to each channel. The F0 estimator was designed to provide an accurate estimate of F0 with minimal processing lag and robustness to the effects of competing noise. Error rates for the F0 estimator and accuracy of the harmonic probability estimator were compared with previous approaches and outcomes demonstrated that the strategy operates effectively across a range of signals and conditions that are relevant to cochlear implant users.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.