Abstract

This study was aimed to establish a standard protocol and to quantitatively assess the reliability of harmonic analysis of the radial pulse wave measured by a harmonic wave analyzer (TD01C system). Both intraobserver and interobserver assessments were conducted to investigate whether the values of harmonics are stable in successive measurements. An intraclass correlation coefficient (ICC) and a Bland–Altman plot were used for this purpose. For the reliability assessments of the intraobserver and the interobserver, 22 subjects (mean age 45 ± 14 years; 14 males and 8 females) were enrolled. The first eleven harmonics of the radial pulse wave presented excellent repeatability (n}{}text {ICCs}>0.9n and n}{}text {p}<0.001n) for the intraobserver assessment and high reproducibility (ICCs range from 0.83 to 0.96 and n}{}text {p}<0.001n) for the interobserver assessment. The Bland–Altman plot indicated that more than 90% of harmonic values fell within two standard deviations of the mean difference. Thus, we concluded that the harmonic analysis of the radial pulse wave using the TD01C system is a feasible and reliable method to assess a hemodynamic characteristic in clinical trial.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.