Abstract

Matrix modifiers have been compared for the determination of cadmium in foodstuffs by ETA—AAS with the sample injected in the form of a slurry. Addition of 800 μg/ml Pd stabilized cadmium to a similar extent as did ammonium dihydrogen phosphate, but avoided the increase in background signal associated with the latter. An analytical procedure was developed, based on palladium matrix modification, platform atomization with a pre-atomization cooling step and integrated absorbance measurements. The method allowed the analysis of milk, liver and olive leaf slurries at concentrations up to at least 50 mg/ml by direct calibration with aqueous standards. The accuracy of the analytical results was within 15% and the detection limit for cadmium in analysis of a 50 mg/ml slurry was 10 ng/g.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.