Abstract

A sensitive and robust trap method was developed for the determination of cadmium (Cd) by using a slotted quartz tube. Using this method at a sample suction rate of 7.4 mL/min for 4.0 min collection, a 1467-fold increase in sensitivity was obtained compared to the flame atomic absorption spectrometry method. Under the optimized conditions, a limit of detection of 0.075 ng mL-1 was obtained for the trap method. The interference effects of hydride-forming elements, transition metals, and some anions on the Cd signal were investigated. The developed method was evaluated by analyzing "Sewage Sludge-industrial origin (BCR no: 146R)", "NIST SRM 1640a Trace elements in natural water", and "DOLT: 5 Dogfish Liver". There was a good agreement between the certified and found values at the 95% confidence level. This method was applied successfully for the determination of Cd in drinking water and some fish tissue samples (liver, muscle, and gill) obtained from Muğla province.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.