Abstract

A sensitive and robust trap method was developed for the determination of cadmium (Cd) by using a slotted quartz tube. Using this method at a sample suction rate of 7.4 mL/min for 4.0 min collection, a 1467-fold increase in sensitivity was obtained compared to the flame atomic absorption spectrometry method. Under the optimized conditions, a limit of detection of 0.075 ng mL-1 was obtained for the trap method. The interference effects of hydride-forming elements, transition metals, and some anions on the Cd signal were investigated. The developed method was evaluated by analyzing "Sewage Sludge-industrial origin (BCR no: 146R)", "NIST SRM 1640a Trace elements in natural water", and "DOLT: 5 Dogfish Liver". There was a good agreement between the certified and found values at the 95% confidence level. This method was applied successfully for the determination of Cd in drinking water and some fish tissue samples (liver, muscle, and gill) obtained from Muğla province.