ABSTRACTHydride generation atomic absorption spectrometry was used as a simple, sensitive, and economical method for the determination of cadmium at trace levels. This study includes a stepwise optimization of all parameters that have an impact on the production, transport, and atomization of cadmium hydride for rapid determination due to the unstable nature of the hydride produced. The cell temperature for atomization, pump period, and flow rate of carrier gas were varied, and the best conditions were defined to obtain lower detection limits. In addition, the acid and sodium borohydride concentrations were optimized since they directly affect the production of the cadmium hydride. Under optimum experimental conditions, the hydride generation efficiency was greatly enhanced to record optimized analytical signals. The accuracy of the method was determined by the analysis of a certified reference material, and the results were in good agreement with the certified value. A linear dynamic range for cadmium was observed from 0.10 to 2.0 µg/L. High sensitivity was obtained with a limit of detection of 0.029 µg/L.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.