Abstract

We have developed a fast and accurate method that uses a small volume of sample to determine over 25 of the typically reported amino acids in human plasma. Samples were prepped with a single step using a spin filter to remove proteins, avoiding the decreased sensitivity from dilution in acid precipitation. Using a reverse phase (RP) High Performance Liquid Chromatography (HPLC) system with O-phthaldehyde (OPA) as the pre-column derivatization reagent, and UV detection at 338 nm, we did a direct comparison with the most common ion exchange/ninhydrin method used in clinical labs on the same plasma samples with 95% concurrence, analysis of amino acid standard solutions returned 99% concurrence. With a sample preparation time of 30 min, utilizing less than 25 μl of sample and with a chromatography run of 30 min, this method can substantially increase access to analysis in both clinical and research laboratories using instruments that are more widely available. SynopsisWe describe a rapid and easily deployed method for sensitive amino measurement in biological samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.