Abstract

Since no proper method is available in literature for the analysis of pyridoxal-5'-phosphate, a reversed phase liquid chromatographic method was developed and validated for specificity, sensitivity, linearity, precision and accuracy. Nine potential related substances and forced degradation products could be successfully separated from the main peak. The separation was achieved on a Polaris C18 column (250×4.6mm i.d., 5μm) using a mobile phase consisting of 20 mM ammonium formate in 0.65% formic acid - acetonitrile (98.8:1.2,v/v). Isocratic elution was performed at a flow rate of 1.0 mL/min and the analytes were detected by UV at 240 nm. The volatile mobile phase allowed also direct coupling to an ion-trap mass spectrometer with a positive electrospray ionization source to characterize unknown peaks in the chromatogram. The method can be used for quality control purposes as required by regulatory authorities to ensure the product's safety and efficacy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.