Abstract

We proposed a Rayleigh-Brillouin scattering (RBS) spectrometer based on a virtually imaged phased array (VIPA) for fast measurements of high-gas temperature. We measured the RBS spectra of air in the temperature range of 374 to 1073 K with an acquisition time of 7 s. We used the Tenti S6 model to fit the spectra and retrieve the absolute temperature values. The root mean square errors of spectra fit residual were less than 3.05%, and the absolute error of the retrieved temperature was less than 39 K. This study demonstrated the ability of the RBS spectrometer to realize fast high-temperature measurement and its potential for combustion control applications.