Abstract

In order to meet the demand of on-site measurement for radionuclides, a portable liquid scintillation TDCR system was developed at National Institute of Metrology (NIM), China. The system consists small size TDCR counter for the measurement of liquid scintillation sources, and digital electronics for pulse signal processing. The optical chamber adopts Teflon material with high diffuse reflection efficiency. Two independent signal processing solutions were used here for TDCR counting. One employed the on-line TDCR solution based on FPGA counting module named TDCR-DMCA, and the other adopted the off-line TDCR solution based on a stand-alone desktop digitizer of CAEN. Two solutions are applied to perform coincidence, dead-time and counting operations follow by MAC3 logic. The performance of the TDCR counting system was tested in benchmark comparison with the traditional custom-built TDCR counting system at NIM through activity measurements of 3H, 14C. Good agreement between these two systems was observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.