Abstract
An interface in elastomeric poly(dimethylsiloxane) (PDMS) for on-line orthogonal coupling of packed capillary liquid chromatography (LC) (i.d. = 0.2 mm) with capillary electrophoresis (CE) in combination with sheathless electrospray ionization (ESI) time-of-flight mass spectrometric (TOFMS) detection is presented. The new interface has a two-level design, which in combination with a continuous CE electrolyte flow through the interface provides integrity of the LC effluent and the CE separation until an injection is desired. The transparent and flexible PDMS material was found to have a number of advantages when combined with fused silica column technology, including ease to follow the process and ease to exchange columns. By combining conventional microscale systems of LC, CE, and ESI-MS, respectively, the time scales of the individual dimensions were harmonized for optimal peak capacity per unit time. The performance of the LC-CE-TOFMS system was evaluated using peptides as model substances. A S/N of about 330 was achieved for leucine-enkephaline from a 0.5 microL LC injection of 25 microg/mL peptide standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.