Abstract
The development of a photothermal absorbance detector for use with microfluidic devices is described. Unlike thermo-optical techniques that rely on measuring refractive index changes, the solution viscosity is probed by continuously monitoring solution conductivity. Platinum electrodes microfabricated on a quartz substrate and bonded to a substrate containing the microchannels enable contact conductivity measurements. The effects of excitation frequency and voltage, electrode spacing, laser power, and laser modulation (chopping) frequency were evaluated experimentally. In the current configuration, a limit of detection of 5 nM for DABSYL-tagged glucosamine was obtained using long injections (to give flat-topped peaks). This corresponds to an absorbance of 4.4 x 10(-7) AU. Separation and detection of DABSYL-tagged glycine, proline, and tryptophan are also shown to demonstrate the feasibility of the method. In addition, simulations were used to investigate the applicability of the technique to small volume platforms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.