Abstract

This paper describes an indigenous method for development and deployment of rechargeable liquid filled phantom with newly proposed radionuclide 141Ce for determination of extrinsic uniformity of gamma cameras. Details about design of phantom, neutron irradiation of cerium targets, chemical processing of 141Ce, charging of phantom with 141Ce solution and their performance evaluation are presented. Suitability of 141Ce in quality assurance of gamma cameras used in in-vivo diagnostic imaging procedures has been amply demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.