Abstract

We created a phantom and analysis program for the assessment of IGRT positional accuracy. We verified the accuracy of analysis and the practicality of this evaluation method at several facilities. End-to-end test was performed using an in-house phantom, and EPID images were acquired after displacement by an arbitrary amount using a micrometer, with after image registration as the reference. The difference between the center of the target and the irradiated field was calculated using our in-house analysis program and commercial software. The end-to-end test was conducted at three facilities, and the IGRT positional accuracy evaluation was verified. The maximum difference between the displacement of the target determined from the EPID image and the arbitrary amount of micrometer displacement was 0.24 mm for the in-house analysis program and 0.30 mm for the commercial software. The maximum difference between the center of the target and the irradiation field on EPID images acquired at the three facilities was 0.97 mm. The proposed evaluation method using our in-house phantom and analysis program can be used for the assessment of IGRT positional accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.