Abstract
The Teledyne Advanced Pollution Instrumentation (TAPI) model 602 BetaPlus particulate system provides nondestructive analysis of particulate matter (PM2.5) mass concentration. This instrument was used to determine if measurements made with cation exchange membranes (CEM) were comparable to standard methods, the β attenuation method at two locations in Reno, NV and an environmental β attenuation method and gravimetric method at Great Basin National Park, NV. TAPI PM2.5 CEM measurements were statistically similar to the other three PM2.5 methods. Once this was established, the second objective, a destructive method for measurement of reactive mercury (RM = gaseous oxidized and particulate bound Hg), was tested. Samples collected at 16.7 L per min (Lpm) for 24 h on CEM from the TAPI were compared to those measured by the University of Nevada, Reno-Reactive Mercury Active System (UNRRMAS, 1 Lpm) CEM and a Tekran 2537/1130/1135 system (7 Lpm). Given the use of CEM in the TAPI and UNRRMAS, we hypothesized that both should collect RM. Due to the high flow rate and different inlets, TAPI data were systematically lower than the UNRRMAS. Correlation between RM concentrations demonstrated that the TAPI may be used to estimate 24 h resolution RM concentrations in Nevada.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.