Abstract

A paper-based colourimetric assay for the detection of alanine transaminase has been developed. In the presence of alanine transaminase, 2,4-dinitrophenyl hydrazine changes to pyruvate hydrazone leading to a colour change from pale yellow to dark yellow. Reaction conditions were optimized using absorption spectroscopic studies. Hydrophobic patterns on the Whatman chromatographic paper were created by wax printing, and the reagents were drop cast at the reagent zone. On the paper device, the intensity of the yellow colour increases with ALT concentration in the range of 20โ€“140 U/L in human serum. For the quantification of ALT, coloured images were captured using a digital camera and were processed with Image J software. The machine learning approach was also explored for the ALT analysis by training with colour images of the paper device and testing using a cross-validation procedure. The results obtained with real clinical samples on the paper device showed good accuracy of less than 5% relative error with the clinical lab results. Furthermore, the paper device shows high selectivity to ALT in the presence of various interfering species in blood serum with a sensitivity of 0.261 a.u/(U/L), a detection limit of 4.12 U/L, and precise results with an RSD of less than 7%. For the testing of whole blood, a plasma separation membrane was integrated with the patterned paper.