Abstract

Okadaic acid (OA), a lipophilic phycotoxin is mainly produced by toxigenic dinoflagellates. The need to develop high performing methods for OA analysis able to improve the traditional ones is evident. In this work, a novel experimental methodology for label-free detection of OA was developed. Protein G magnetic beads (protein-G-MBs) modified gold electrode was used to immobilize anti-OA monoclonal antibody (anti-OA-MAb). Preliminary, colorimetric tests were performed in order to validate protein-G-MBs and anti-OA-MAb reaction. Electrochemical detection was carried out by differential pulse voltammetry in ferri/ferrocyanide solution. The limit of detection value obtained (0.5μgL−1) validated the developed electrochemical immunosensor as a promising tool for routine use. The matrix effect and the recovery rate were also assessed with real samples, showing a good percentage of recovery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.