Abstract

A novel UPLC-UV method was developed for analysis of timolol in human plasma using a simple, fast, and cost effective ion-exchange SPE procedure, followed by separation on a C18 UPLC column with a mobile phase consisting of acetonitrile, phosphate buffer, and sodium 1-octane sulfonate as an ion pairing agent. The method was fully validated according to US-FDA guidelines, and was found to be sufficiently accurate and precise for analysis of timolol in human plasma for clinical pharmacokinetic studies. The application of ion-exchange SPE cartridges for purification of timolol in plasma produced excellent percent recoveries and good sample clean-up, while the ion-pairing separation described here allowed quantitation of timolol without interference from endogenous sample components. The method lower limit of detection was 1.7 ng/mL and the lower limit of quantitation was 5.0 ng/mL, allowing for analysis of therapeutic concentrations of timolol in plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.