Abstract
We demonstrate an optical instrument that can measure the axial displacement of different eye tissues, including the cornea and the fundus. The instrument is based on spectral-domain low-coherence interferometry, which extracts displacement information from sequential axial scans of the eye with 100 Hz sampling frequency and with a precision of 400 nm. Longitudinal retinal and corneal movements were successfully measured in vivo in live rats, and Fourier analysis of the signal revealed the signature of the respiratory and cardiac cycles at 1.0 and 3.5 Hz, respectively. The effective amplitudes of retinal and corneal displacements at the cardiac frequency were found to be about 1.10 and 1.37 μm, respectively. The synchrony and direction of these two movements relative to the systole and diastole were found to be nearly the same. This novel instrument can be applied to assess biomechanical properties of the eye, which could be important for early diagnosis and for understanding the pathophysiology of glaucoma and other ocular diseases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.