Abstract

Although the assessment of α-amylase is an essential part of the diagnostic workout of several pancreatic and extra-pancreatic disorders, its enzymatic activity is significantly reduced in the presence of cell-free hemoglobin such as in samples with spurious hemolysis, due to chemical and spectrophotometric interference. We developed a new reagent that provides reliable results on hemolyzed biological specimens. All tests were performed on Beckman Coulter AU5822. Intra-assay imprecision was assessed on three serum samples with low, intermediate and high α-amylase concentration. Linearity was assessed by serially diluting two samples with low and high values of α-amylase. The comparison with commercial reagent was performed on 40 serum samples. Hemolysis studies were carried out by mechanically hemolysis of 20 lithium-heparin samples. The intra-assay imprecision was comprised between 1.3% and 2.2%. The linearity was excellent (r=0.998), and highly significant agreement was observed with the commercial assay (r=1.00; mean bias -3.8%). Although a significant correlation between non-hemolyzed and hemolyzed specimens was found with both assays (p<0.001), a much greater agreement was observed with the experimental method (r=0.997 vs. 0.818). No measurement exceeded the total allowable error with the experimental assay, whereas the threshold was exceeded in 85% of samples using the commercial method. The clinical applications of the experimental reagent include α-amylase assessment in hemolyzed samples, in urine and other biological fluids contaminated with lysed erythrocytes, or in patients under frequent transfusions and hemoglobin-based blood substitutes therapy. The formulation of this reagent could be adapted for other clinical chemistry or immunochemistry assays.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call