Abstract
A new, simple and rapid method for the quantitative determination of the antimicrobial preservative 2-phenoxyethanol, based on reverse phase ultra-high-performance liquid chromatography has been developed. The validation was performed according the ICH Q2 guideline “Validation of Analytical Procedures”. The desired chromatographic separation was achieved on a Waters Symmetry C18 (150 × 4.6 mm, 5 μm) column using an isocratic elution, with detection at 270 nm wavelength. The mobile phase consisted of acetonitrile/water (55:45, v/v), pumped at a flow rate of 1 mL/min. The calibration curve and the analytical procedure are linear (r2 = 0.999) from the concentration of 0.07 mg/mL to 1.1 mg/mL. The percent relative standard deviation for intra- and inter-day precision was <1%. The recovery of 2-phenoxyethanol in vaccines ranged between 96.5 and 100.60%. The limits of detection and quantitation were 1.3 × 10−4 and 2.7 × 10−4 mg/mL, respectively. The method was found to be robust by changing the column working temperature, the percentage of acetonitrile of the mobile phase and the flow rate. The validated method can be successfully and reliably used to quantify as well as to exclude presence of 2-phenoxyethanol preservative in marketed vaccines.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.